Oxygen concentration: Sensors

Appropriate sensor is fundamental to any measurement setup. Variety of different types and variants are available, depending on the measured quantities, sensor dimensions, as well as usage workflow, environment, and installation method. The vast majority of sensors listed below are just components of a larger measurement setup, and can therefore be satisfactorily used only with a compatible analyzer. Sensors for in-technology measurements should be installed inside our probes, ensuring their fixing and protection, and these connect to a terminal block. Variants for laboratory and portable usage are terminated with the appropriate connector. Sensors generally have the shortest lifetime of the entire setup, due to direct contact with the measured medium.

Optical oxygen sensors CSOT53

Photo

CSOT53 sensors are mainly used to measure and control the oxygen content in the activation tanks of biological WWTPs, in surface, underground and waste water or in fish farming facilities. CSOT53 is an optical sensor based on the principle of luminescence quenching by oxygen, fitting into a classic miniature outline (no input block integrated).

The P versions are designed for installation into fixed probes, while the L versions are suitable for use with desktop and portable instruments. We also offer a version compatible with standardized ground glass joints (primarily for BOD measurement).

  • Accurate measurement in both still and flowing water, thanks to zero dependency on flow rate.
  • Smooth surface of the plastic membrane allows self-cleaning of the sensor in flowing water (typical for WWTP activation tanks) and thus lower maintenance requirements.
  • Long service life of the sensor and the membrane head, with maximum simplicity of head replacement.
  • Virtually no need for recalibration.
  • Resistance to the effects of CO2, SO2, H2S significantly increases the reliability of measurements, especially in WWTPs.
TypeSuitable forDimensionsOperating and storage temperature
CSOT53PSInstallation with SPO41, SPO41K, SPO42, SPO41MEK, PB42V, PB43V, PE G3/4KØ14.5 x 57 mm-5 to +50 °C
CSOT53PLInstallation with SPR42, SPR41ME, PE G3/4L, SVK42Ø14.5 x 113 mm-5 to +50 °C
CSOT53PSNInstallation with SPO41MEK, PE G3/4KØ14.5 x 57 mm-5 to +70 °C
CSOT53SInstallation inside pipes (stainless steel fitting)Ø14.5 / Ø25 x 127 mm-5 to +70 °C
CSOT53LSLaboratory measurementsØ14.5 / Ø18 x 158 mm-5 to +50 °C
CSOT53LZPortable instruments (with protective cover)Ø14.5 / Ø18 x 98 mm-5 to +50 °C
CSOT53LPSLaboratory, measurements in enclosed unitsØ14.5 x 79 mm-5 to +50 °C
CSOT53LLLaboratory, measurements in bottles (extended length)Ø14.5 / Ø18 x 208 mm-5 to +50 °C
CSOT53LSTLaboratory, for containers with ground glass jointsØ14.5 x 108 mm, ST 19/26-5 to +50 °C
PrincipleOptical, quenching of luminescence by oxygen
CompatibilityINSA analyzers and instruments only
Measurement range0.1 - 20 mg/l or 1 - 150 %
Resolution0.01 mg/l or 0.1 %
Response timeca 45 sec.
Flow of the sampleNot required
Pressuremax. 0.6 MPa
Materials in contact with mediumPolyethylene terephthalate, EPDM, PTFE, PP, silicone, stainless steel

Oxygen sensors CSOT43

Photo

CSOT43 sensors are mainly used to measure and control the oxygen content in the activation tanks of biological WWTPs, in surface, underground and waste water or in fish farming facilities. CSOT43 is a classic Clark type sensor, with robust all-glass two-electrode detection system and a small-area cathode. Compared to modern optical sensors, its advantage is primarily a faster response, as well as the ability to measure very high values.

The P versions are designed for installation into fixed probes, while the L versions are suitable for use with desktop and portable instruments.

  • Optimal size of the electrodes and the electrolyte supply ensures perfect linearity, excellent time stability and long maintenance-free operation, without the need for electrolyte replacement, or a third electrode.
  • Extremely low oxygen consumption reduces the demands on the flow of the measured water. The sensor can work in virtually any conditions, without additional stirring.
  • Smooth surface of the plastic membrane allows self-cleaning of the sensor in flowing water (typical for WWTP activation tanks) and thus lower maintenance requirements.
TypeSuitable forDimensionsOperating and storage temperature
CSOT43PSInstallation with SPO41, SPO41K, SPO42, SPO41MEK, PB42V, PB43V, PE G3/4KØ14.5 x 57 mm-5 to +40 °C
CSOT43PLInstallation with SPR42, SPR41ME, PE G3/4L, SVK42Ø14.5 x 113 mm-5 to +40 °C
CSOT43PSNInstallation with SPO41MEK, PE G3/4KØ14.5 x 57 mm-5 to +80 °C
CSOT43SInstallation inside pipes (stainless steel fitting)Ø14.5 / Ø25 x 127 mm-5 to +80 °C
CSOT43LSLaboratory measurementsØ14.5 / Ø18 x 158 mm-5 to +40 °C
CSOT43LZPortable instruments (with protective cover)Ø14.5 / Ø18 x 98 mm-5 to +40 °C
CSOT43LPSLaboratory, measurements in enclosed unitsØ14.5 x 79 mm-5 to +40 °C
CSOT43LLLaboratory, measurements in bottles (extended length)Ø14.5 / Ø18 x 208 mm-5 to +40 °C
PrincipleClark type sensor
Measurement range0.1 - 30 mg/l
Electrode currentca 10-8 A (open air, 25°C), ca 10-10 A (N2)
Response timeca 15 sec.
Flow of the samplemin. 2 mm/s (for 1 % output tolerance)
Materials in contact with mediumPolyethylene terephthalate, PTFE, PP, silicone, stainless steel