Chlorine dioxide concentration: On-line analyzer MFD88
The MFD88 is a device intended for fixed installation in the industry. It supports up to 12 measuring points in any combination of the measurement types offered. The outputs are provided through current loops, output relays and a digital interface, suitable both for connection with usual control systems and for direct control of smaller technologies.
Measurement locations are equipped with so-called input blocks located near the sensors. This allows a distance of up to 1000 m between the sensors and the device, without affecting the reliability of the measurement. A single unshielded two-core cable with the possibility of branching is sufficient to connect all the measured points. Housing for sensors and input blocks is provided by a wide variety of different probes, designed to suit sites of different nature. The device also supports automatic sensor cleaning, temperature correction and supplementary temperature measurement for each location.
Output interfaces include four industry standard 0-20 / 4-20 mA analog current loops with configurable ranges, up to 4 relays (expandable to 8 using current loops), as well as a MODBUS / RS485 digital interface. It is thus possible to directly connect to common current loop based configurations, to directly control technological elements (such as various dosing valves, etc.), or just as easily to transfer the entire set of measured values to the control system via a single digital interface.
Interaction with the user is easy thanks to the high-contrast dot-matrix LCD display, which is easy to read in the dark and in the sun, and a simple menu-based system controlled by only 6 buttons. Easy-to-open transparent door and optional access codes provide a good balance between convenient operation and protection against harsh environments and interventions of unauthorized persons. Semi-automatic calibration procedures are useful for regular maintenance of the measuring set-up.
The installation options are quite wide. Thanks to a wide range of permitted temperatures, a high level of protection, an alternative power supply of 230V AC or 24V DC, a large permitted length of cabling and overvoltage protection applied to all interfaces (against the secondary effects of lightning), there is no problem placing the device almost anywhere. An additional protective cover is available for outdoor use (the MUT13V clamp).
Advanced functions include freezing of outputs during calibrations or sensor cleaning, to prevent interference with technology control during measurement downtime, data logging available for later review or download, storage of calibration data in the input block located in the probe, allowing easy relocation of the probe for maintenance or other use without loss of calibration, as well as a number of options for direct control of the technology via relays or current loops, with limiting values control, time delay and other widely configurable logic, or using integrated PID controllers with convenient automated settings.
The order must include a list of the required measurements and additional components, so that appropriate probes equipped with the correct sensors and input blocks can be supplied, as well as the correct device configuration for the setup.
Chlorine dioxide measurement
Chlorine dioxide measurement works with CSDT43 sensors. The measurement is suitable for continuous monitoring and regulation of the chlorine dioxide concentration, but must be regularly calibrated (usually by the DPD method). Where necessary, the same device may also process the measurements of pH or other quantities.
Automatic sensor cleaning
Automatic sensor cleaning may be used on a limited number of measuring points, if they are equipped with SPO41ME or SPR41ME probes and sensors suitable for automatic cleaning. The device controls the mechanisms of the individual probes, and freezes the respective values so that the output from the device is not affected by the ongoing cleaning process.
Power supply | 230 V 50 Hz, or 24 V DC ± 20 % |
Power consumption | max. 20 VA |
Dimensions (cable glands not included) | 240 x 184 x 115 mm |
Weight | ca 1.5 kg |
Enclosure | IP 65 |
Operating temperature | -20 to +50 °C, input block -20 to +70 °C |
Relative humidity | 10 to 90 % |
Analog outputs | 4x loops 0 - 20 or 4 - 20 mA, configurable transfer from measurement display |
Analog outputs load | Passive, max. 500 Ω |
Relays | 2x or 4x built-in relay (further relays 12V/20mA may be connected to analog outputs) |
Contact rating (built-in relays) | max. 3 A at 230V/50Hz, max. 5 A at 24V DC |
Temperature measurement range | -2 to +100 °C (also limited by installed sensor) |
Temperature measurement accuracy | ± 0.2 °C |
Chlorine dioxide measurement range | 0.01 to 2 mg/l |
Chlorine dioxide measurement accuracy | ± 5 % (of full range, at constant temperature) |
Automatic cleaning | max. 4 measuring points |
Automatic cleaning interval | Configurable from 1 min. to 24 hrs. |